How to operate the hottest ring press

2022-08-24
  • Detail

How to operate the ring press

hd-a ring press is one of the most commonly used products in the packaging industry. It serves Guangdong, Guangxi, Hainan and other regions, and is also the most sold product of Haida. It is found that many customers are not very familiar with the operation after purchasing the ring press. The following Haida technician will introduce you to the operation method of the ring press:

I Prepare samples:

1. Determination of sample position: select non scar samples from the dispersion position of the same batch of materials

2. Sample making: cut samples with a width of 12.7 0.1 mm and a length of 152.0 0 0.2 mm

3. Number of samples: take more than 10 pieces in the vertical and horizontal directions

4. Standard state of the sample: before the test, it shall be placed in the state of temperature 23.2 ℃ and relative humidity 60.5% for more than 24 hours

note

n

g

use gt-hy ring pressure tester to cut the standard sample (purchased separately)

n the edge of the slice shall not have burrs or other defects that affect the measurement results

n hand contact with gloves is required in the test

n the specimen whose long side is parallel to the longitudinal direction is used to measure the transverse hoop compression strength

n the specimen whose long side is parallel to the longitudinal direction is used to measure the transverse hoop compression strength

II Setting of upper and lower safety protection devices:

1. The functions of the upper and lower limit devices are mainly:

a to prevent the upper pressure plate from falling to impact the lower pressure plate or the test flat fixture, and damage the load element

b as long as the touch piece of the upper pressure plate touches the upper limit setting button or the lower limit setting button, the upper pressure plate will stop automatically

c the distance between the upper and lower pressure plates or the upper pressure plate and the test fixture can be adjusted

2. First fix the upper and lower limit fixing blocks at the appropriate position

3. The upper limit setting button should be set at a position sufficient for the upper pressure plate to place the test fixture or sample of the device, and the lower limit setting button should be set at a position where the upper pressure plate drops to the upper and lower pressure 5 gold tool torque detection equipment parameters: the plate or the upper pressure plate and the test fixture do not collide with each other

4. Loosen the screw of the upper or lower limit setting button to move the setting button When moving to positioning, it must be tightened on the touch rod

warning

n

the upper and lower limit fixing blocks have been adjusted before leaving the factory. Do not move them unless necessary

n after replacing different test fixtures, the position of the upper and lower limit setting buttons must be reset

n when locking the upper and lower limit setting buttons, it should be locked firmly to prevent slipping and crashing the load element

n after setting, test whether the function of the upper and lower limit setting button protection device is normal

III. operation steps

1. Select the diameter of the inner disc according to the thickness of the sample and place it on the sample seat (the thickness of the inner disc is marked as mm), carefully insert the sample into the sample seat and ensure that it is inserted to the bottom

3. The machine must be connected to 220V 5% 50Hz power supply, and good grounding protection measures should be taken

4. Turn on the power switch on the panel, the panel will be powered on, and the display will automatically open the program

5. confirm whether the travel protection device of the machine is effective and whether the position is reasonable

6. press the reset key to set to zero, then press the peak key, press the down key, and press the test key to test after reaching the surface of the required test object

7. place the test sample in the middle of the pressing plate, press the zero key and the peak key, and continue to press the down key to start the test

8. it is strictly forbidden to directly contact the upper and lower pressing plates or test with metal specimens

9. during the test, if there is any abnormal condition, press the stop key and immediately notify the maintenance personnel for detection. It can only be used after troubleshooting

10. the display receives the signal from the pressure sensor on the top of the upper pressing plate and displays the current pressure value. Press the zero key to clear the numbers on the

display. Press the peak key to record the maximum value during the test

11. date, time, breakpoint ratio and constant value setting parameter setting:

step

operation

display

description

1

after power on

press [function]

[D 060311]

after power on, press and hold [function] and hold

2

press [function]

Press [Print]

Press [peak]

[D 060322]

date setting

if it needs to be modified, press [function] to shift, and press [Print] to increase the number Press the

[peak] key to confirm

3

press [function]

Press [Print]

Press [peak]

[t 101208]

time setting

if it needs to be modified, press [function] to shift, and press [Print] to increase the number Press

[peak] key to confirm

4

press [function]

press [Print]

press [peak]

[b 85]

breakpoint. In addition, control the teeth at about 60 degrees or set the ratio of the mouth smaller

if it needs to be modified, press [function] to shift, and press [Print] to increase the number Press

[peak] key to confirm

5

the following is an introduction to the relevant functions of the pellet pressure testing machine. Jinan new era Gold Testing Instrument Co., Ltd. can be used as a reference for users and friends: [function]

Press [Print]

Press [peak]

[a 50.00]

quantitative value setting

if it needs to be modified, press [function] to shift, and press [Print] to increase the number Press the

[peak] key to confirm the

Article source: Dongguan Haida international instrument is prohibited to reprint without permission! (or: if you need to reprint, indicate the source)